Full Auto Dry Biochemistry Analyzer

Band nameICLEAR
Model No.CHL-D100
CertificateISO CE
Add to Favorites

Quick Details

Specification
Sample type: Anti-coagulation whole blood, serum, plasma. Sample volume: 100ul.
Bar code: Two-dimensional bar code.
Testing time :12mini/person
Testing principle : absorption spectroscopy, transmission turbidimetry Testing method: end point, kinetic, fixed time, turbidmetry etc. Temperature: 37°C±0.2°C.
Absorbance :0-3.0Abs
Resolution: 0.001Abs.
Carryover: 0
QC & Calibrate: auto test in real time.
Work environment:
Temperature:0-35°C Humidity:<90%
Optic system: Reversed optic. 340nm, 405nm, 450m, 510nm, 546nm, 578nm, 630nm, 850nm.
Light source: 10W Stroboscopic xenon lamp, ultra-long life.
Power supply:AC100V-240V,50-60Hz;DC:9~15V
Power: 50W
7.0 inch True color LCD touch screen,multi-language.
Vet version is optional.
Storage: internal storage 64M, >100000 results.
Printer: built-in thermal printer.
Data port: 4 USB, 1 RS232.
Weight:5KG
Size: 31cm(H)*21cm(W)*28(L)cm 

Packaging & Delivery

5KG    Size: 31cm(H)*21cm(W)*28(L)cm

Product Description

Full Auto Dry Biochemistry Analyzer
CHL-D100


Advantage 

Simple operation
Free of centrifugation, water treatment system and professional doctor. Anyone can operate only by 3 steps. Add 100ul whole blood into disc, put disc into analyzer and delay 12 minutes to print report. 
Little sample volume
Sample volume is 10-20% of normal chemistry analyzer.
Accurate Result
Test sample and dilution volume measured by mould, without random error. No carryover.
Portable,5 Kg, 0.02m3
Provide with AC and DC power, suitable for use at movable place and outdoor. Free maintenance
Auto QC and real time Calibrate.
Stroboscopic xenon lamp, service life is over one hundred million times. 
No internal fluid flow, no pump and valve, decrease machine fault. 
Specification
Sample type: Anti-coagulation whole blood, serum, plasma. Sample volume: 100ul.
Bar code: Two-dimensional bar code.
Testing time :12mini/person
Testing principle : absorption spectroscopy, transmission turbidimetry Testing method: end point, kinetic, fixed time, turbidmetry etc. Temperature: 37°C±0.2°C.
Absorbance :0-3.0Abs
Resolution: 0.001Abs.
Carryover: 0
QC & Calibrate: auto test in real time.

Work environment:
Temperature:0-35°C Humidity:<90%
Optic system: Reversed optic. 340nm, 405nm, 450m, 510nm, 546nm, 578nm, 630nm, 850nm.
Light source: 10W Stroboscopic xenon lamp, ultra-long life.
Power supply:AC100V-240V,50-60Hz;DC:9~15V
Power: 50W
7.0 inch True color LCD touch screen,multi-language.
Vet version is optional.
Storage: internal storage 64M, >100000 results.
Printer: built-in thermal printer.
Data port: 4 USB, 1 RS232.
Weight:5KG
Size: 31cm(H)*21cm(W)*28(L)cm 

Advantage of disposable Reagent disc
1. Freeze-drying reagent, pack per person, storage life as long as 1 year.
2. Internal autonomous channel design, disposable disc avoid carryover. 3.Multi reagent disc for choose. 

Introduction of Reagent Disc
14Conventional Parameters
TP,TC,TG,GLU,TBIL,Crea,UREA,UA,AMY,ALT,AST,GGT,ALP,ALB,GLOB,ALB/ GLOB 

11 Hepatic Function
TP,ALB,AMY,TBIL,ALT,GGT,ALP ,AST,TC,CHE,TBA

8 Kidney function
GLU),ALB,Crea,UREA,UA,K+,Na+,Cl-,Ca2+,CO2,P 

7 electrolyte
K+,Na+,Cl-,Ca2+,CO2,Mg2+,P 

6 Lipid Parameters
GLU,TC,TG,ALT,AST,HDL,TC/HDL,LDL,VLDL 

15 comprehensive metabolism
TP,GLU,TBIL,Crea,UREA,UA,ALT,AST,ALP,ALB,K+,Na+,Cl-,Ca2+,CO2,GLOB,ALB/ GLOB 