900T/H Fully Automatic Biochemical Analyzer

Band nameICLEAR
Model No.CHL-G1200
CertificateISO CE
Add to Favorites

Quick Details

● Fully automatic discrete with emergency function
● 900 tests / hour reagent plate(600 colorimetry + 300 electrolyte)90 positions
● 115 sample bits 120 cuvettes
● 24 hours ready to use

Packaging & Delivery

1220mm*800mm*1150mm 230Kg

Product Description

900T/H Fully Automatic Biochemical Analyzer

CHL-G1200

Type of System
● Automated discrete, STAT priority
 
Throughput
● Biochemistry 600 test/h, ISE 300 test/h
 
 
 
Measuring principle
● Endpoint, rate, fixed-time
 
● Colorimetry,turbidimetry,ion-selective electrode technology
 
 
Sample handling
● 115 sample positions,continuous loading,support polytype primary tube& microcontainer cup
 
Reagent handling
● 2 disks, 90 positions,refrigerated with independent power switch, continuous loading
 
Sample volume
● 2 μL-35μL, 0.1μL stepping
 
Reagent volume
● 20 μL-350μL,1μL stepping
 
Reaction volume
● Minimal reaction volume as low as 150 μL
 
Reaction cuvette
● 120,optical diameter is 6 mm
 
Reaction temperature
● 37℃±0.1℃
 
Light source
● Halogen lamp, lifespan up to 2500 hours
 
Wavelength
● 340nm,380nm,405nm,480nm,505nm,546nm,570nm,600nm,630nm,660nm,
700nm,720nm,750nm,780nm,800nm
 
Water consumption
● Deionized water 16 liter/hour