Fully Automatic Chemistry Analyzer
CHL-GS300
System function
Throughput: up to 225 tests/Hour(not including ISE tests)
Test Method: End point, Fixed time, Kinetic
Test Sequence: Discrete, Random access, STAT can be inserted at any time. Test sequence can be
sorted by samples or items
Spectrometer: enclosed array of spectrometer and photodiodes, maintenance free.
Cuvette washing: auto wash the cuvettes through 8 segments washing pipe and continuous use them(optional)
Reaction detect: monitor 81 cuvettes’s reaction at the same time
Photometric process: the interval time between two consecutive photometric points are exactly the
same.
Pipeline system: pressure auto released pumps and double filters to protect pumps and valves
Gearing: plastic gear leads lower noise and accuracy of positioning of moving parts
Environmental resistance: resistant form dust and water vapor
Sample/Reagent/Stirrer system
Sample volume:2μL~30μL,step by 0.1μL
Sample disc: 60 positions for primary tubes and secondary cups
Reagent volume:20μL~300μL, step by 1μL
Reagent disc: provide 60 reagent bottle positions
Reagent bottle: particular reagent bottle without dead volume
Reagent Cooling: continuous cooling system according to the principle of peltier
Probe: inner and outer walls are highly polished leads carry-over <0.1%, level detection and
vertical and horizontal collision protection
Stirrer: paddle stirrer coated with Teflon, stir immediately after the sample was added in.
Water consume: <1.5L/hour
Carry over: Optimized item test sequence and reinforced washing process, lower the carry over and
avoid probe being clotted
Reaction system
Cuvette: permanent UV free glass cuvette
Reaction disc: Discal reaction disc with 81 disposable cuvettes
Reaction volume:150μL~600μL
Heater: solid and sealed heater leads temperature raise rapidly and steady and maintenance free
Temperature: 37℃±0.1℃
Optical system
Light source: 12V 20W, halogen lamp with lifetime more than 2000 hours
Spectrometer: post spectrometer with arrayed and sealed filters, maintenance free
Wavelength: 340nm、405nm、450nm、510nm、546nm、578nm、630nm、670nm
Absorbance range: 0~6
Calibration/QC
Calibration type: single point linear、two points linear、multi-points linear、Logit-4P、Logit-5P、
Spline、Exponential 5P、Polynomial 5P, auto choose the best fittest calibration
curve.
QC type: real time QC、daily QC、day to day QC
QC chart: L-J、Cumulative sum check、Twin Plot
QC rule: Customized QC rule in order to adapt different item’s requirements
Operational system
Operational system: VISTA、Windows XP、Windows2000
Memory: auto save and recover the results
Report format: Customized report format
Status Check: real time check the working status of the analyzer
Power/ Dimension
Voltage: AC 85V~264V, 55Hz±8Hz, Auto adapt according to the input voltage
Power Consume: ≤350VA
Dimension: 850mm×650mm×1067mm(L×W×H)
Net weight: 90Kg