Lab Electrolyte Analyzer Machine

Band nameICLEAR
Model No.CHL-GE300
CertificateIOS CE
Add to Favorites

Quick Details

Throughput: 60 samples per hour
Sample volume: 100-200μL
Ion: K+, Na+, Cl-, iCa2+, pH, Li+, customized selection of ions, optional CO2
Calibration: Automatic or On-demand
Closed reagent, real time monitoring reagent
Printer: Internal thermal printer
Display: LCD
Storage: Up to 50,000 results
Interface: RS-232, support LIS

Packaging & Delivery

Dimension 355mm×330mm×175mm 

Net weight 5.6 KG

Product Description

Lab Electrolyte Analyzer Machine

CHL-GE300


General Throughput 60 samples/hour
Principle Impedance for WBC,RBC,PLT; SFT method for HGB
Channels 2
Aperture diatmeter WBC chamber:100um, RBC/PLT chamber:70um
Parameters 20 parameters(including WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%,     
RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR)    
Histograms 3 histograms(including WBC,RBC,PLT histograms)
HGB lamp LED 540nm
Sample volume Normal: 10ul, Prediluted: 20ul
Software Calibration Manual & auto calibration
Counting modes Venous, capilary, prediluted
Print  Auto print, manual print
LIS Support LIS
Language English, Spanish,Italian,Portuguese, etc
Reagent Reagent Standard: Dilent: 20L, Lyse: 500ml
Cleanser Standard: Probe cleanser: 50ml
Specifications Input 10.4 inch touch screen, keyboard, mouse
Output Internal printer, support external printer
Printer paper 80×35mm
Interface RS232 port, USB port, VGA port
Storage 30,000 results with histogram
Dimension 514mm×594mm×645mm
Blockage clear High voltage, high pressure flush
Net weight 24 KG
Others Work temperature 18-35°C
Power requirement 100~240 AC, 50/60Hz
Relative humidity ≤85﹪
Atmospheric pressure 86.0kPa-106.0kPa 
Performance Carry-over rate Parameter CV
WBC ≤ 1%
RBC ≤ 1%
HGB ≤ 1%
PLT ≤ 1.5%
Background Parameter Background
WBC ≤ 0.2×109/L
RBC ≤ 0.02×1012/L
HGB ≤ 1 g/L
HCT ≤ 0.5%
PLT ≤ 5×109/L
Precision Parameter CV Measurement range
WBC ≤ 2.0% (4.0-15.0×109)/L
RBC ≤ 1.5% (3.5-6.0×1012)/L
HGB ≤ 1.5% (110.0-180.0g)/L
MCV ≤ 0.4% (80.0-110.0)fL
PLT ≤ 4.0% (100.0-500.0×109)/L
Linearity Parameter Measurement range CV
WBC (0-5.9×109)/L
(6.0-99.9×109)/L
(±0.3×109)/L
±5%
RBC (0-0.99×1012)/L
(1.0-9.99×1012)/L
(±0.05×1012)/L
±5%
HGB (0-99.0g)/L
(100.0-300.0g)/L
(±2.0g)/L
±2%
PLT (0-99×109)/L
100-999×109)/L
(±8.0×109)/L
±10%