Display 10-inch touch screen
Light source Quartz-halogen lamp
Wavelength range 200 - 1000 nm with 1 nm steps
Wavelength Accuracy ±2 nm
Repetition ±0.2 nm
Wavelength selection Monochromator
200 - 1000 nm 96 and 384-well Plates UV/Vis Microplate Spectrophotometer ELISA Reader
CHL-A200
CHL-A200 is a UV/Vis microplate spectrophotometer offering free wavelength selection, so it is an ideal tool for virtually any photometric research application, such as DNA, RNA, protein analysis and more. It offers exceptional usability through its intuitive, user-friendly internal software and comprehensive ReaderIt-II Software. A200 microplate reader has been designed to deliver high performance and high-quality results with minimal user effort, It measures 96 and 384 well plates and plate shaking motion is linear.
Features:
1. 10-inch touch screen, and offers stand-alone operation, it is easy to use
Intuitive, user-friendly internal software on a 10-inch touch screen for quick measurements. Easy and logical assays are set up for demanding assays with the powerful PC software (ReaderIt-II), the PC software can offer to analyze, review, and validate data. It is easy to access data via the USB interface and WIFI
2. Freely selectable wavelengths from 200 - 1000 nm for the demands of various assays
Wavelength selection in A200 is done by using an advanced monochromator system. Any wavelength between 200 to 1000nm can be selected at any time. Using the spectral scanning feature, the whole spectrum of a sample is scanned in 1nm increments to allow the identification of the optimal measurement wavelength for a new assay.
3. Both microplate readings for any throughput requirements
It measures 96-and 384-well plates
4. Flexibility and performance for various photometric applications
A200 microplate reader supports endpoint, kinetic and spectral scanning assays. It reads 96- and 384-well microplates with and without lids. A broad wavelength range with the UV area, pathlength correction, as well as fast spectral scanning, makes it an ideal tool for any photometric research application, including DNA, RNA and protein analysis
Specifications
Model |
A200 |
Display |
10-inch touch screen |
Light source |
Quartz-halogen lamp |
Wavelength range |
200 - 1000 nm with 1 nm steps |
Wavelength Accuracy |
±2 nm |
Repetition |
±0.2 nm |
Wavelength selection |
Monochromator |
Read-out range |
0 - 4.0 OD |
bottle |
3 pcs 2.5L washing bottle, 1pcs 2.5L waste bottle |
Half-bandwidth of filters |
<2.5 nm |
Detector |
Two silicon photoelectric detectors, one for measurement, |
and another for reference |
Linearity @450nm |
0 - 2.5 Abs,±2% (96 well plates) |
Accuracy @450nm |
1.0% + 0.005 Abs (0 - 2.0 Abs);2.0% [2.0 - 2.5 Abs) |
Precision @450nm |
CV < 0.5% accurate model;CV < 1.0% fast mode |
Test speed |
10 seconds with 96-well plate (fast mode); 30 seconds with 96-well plate (accurate model) |
Plate shaking |
Linear; three options for speed |
Test speed |
15 seconds with a 96-well plate (single wavelength); |
30 seconds with a 96-well plate (double wavelength) |
Incubator |
From ambient + 5℃to 45℃ |
UI |
Integrated software or PC control software |
Analysis software |
ReaderIt-II |
User interface |
Touch screen, android system, 10-inch touch screen, |
external keyboard mouse |
Storage |
16G memory, more than 10,000 test records can be stored |
Ports |
3 USB ports, for PC, printer and USB-disk |
Automated systems |
Temporarily unable compatible with automated systems |
Power supply |
DC24V 6.5A |
Dimensions(W×D×H) |
300×500×290mm |
Weight |
15kg |